Module errors, Media-related errors, Communication errors – Yokogawa PC-Based MX100 User Manual

Page 74

Advertising
background image

3-2

IM MX100-01E

Module Errors

The left and right digits of the 7-segment LED display “E” and an error code,
respectively. The LED illuminates. In the case of module errors, the error number and the
corresponding module number are displayed alternately as shown in the figure below.

Error number

Module number

Display Probable Cause

Corrective Action

Ref. Page

E0

Range information error.

Servicing required.

-

E1

Calibration value error.

Check the module’s installation status, then recalibrate the

module.

If the error occurs even after recalibrating, servicing is required.

*

E2

Calibration reference voltage value is

not correct (during calibration).

Check whether the correct calibration reference voltage is being

applied or whether the channel to which the voltage is applied is

correct.

*

E3

Error in writing the calibration value.

Servicing required.

-

E4

The installed module cannot be used. Replace the module with one that can be used.

-

E5

Initial balancing Error

(Initial unbalance value adjustment)

Check the module’s installation status, then perform initial

balancing again.

If the error occurs even after recalibrating, servicing is required.

-

*

See the MX100 Standard Software User’s Manual (IM MX180-01E).

Media-Related Errors

The left and right digits of the 7-segment LED display “P” and an error code, respectively.
The LED blinks. The error code is saved to the log file (see “Log File Storage” under
“Saving Data to the CF Card” in section 1.2).

Display Probable Cause

Corrective Action

Ref. Page

P0

Manipulated the CF card while it was

being accessed.

Do not manipulate the CF card while it is being accessed.

2-27

P1

CF card error.

The CF card may not be formatted or may be damaged. Check

the CF card.

-

P2

The CF card was not formatted

correctly.

Insert the CF card into the CF card slot and format the card again,

or replace with another CF card.

*

P3

Not enough free space on the CF

card.

Delete unneeded data to increase the free space, or replace with

another CF card.

*

P4

The number of files saved on the CF

card exceeded 512.

Delete unneeded files to decrease the number of files, or replace

with another CF card.

*

P5

Overwrite error in the data storage

buffer.

The time needed to store the file must be reduced. Delete

unneeded data to increase the free space, or format the CF card.
If you are using a software application through the MX100/

DARWIN API, the timeout setting may not be appropriate. If so,

change the value to an appropriate value.

*

*

See the MX100 Standard Software User’s Manual (IM MX180-01E).

Communication Errors

The left and right digits of the 7-segment LED display “C” and an error code, respectively.
The LED blinks.

Display Probable Cause

Corrective Action

Ref. Page

C1

Multiple connection error.

Disconnect the current connection and reconnect. However, this

error may also appear when the Ethernet cable is removed and

inserted again. Turn OFF the power and reconnect.

*

*

See the MX100 Standard Software User’s Manual (IM MX180-01E).

3.1 Error Display on the 7-Segment LED and Corrective Actions

Advertising