Overview of setup pages in cyberscan con 1500 – Thermo Fisher Scientific CyberScan CON 1500 User Manual

Page 13

Advertising
background image

Instruction Manual CyberScan CON 1500

9

4.2.

Overview of Setup Pages in CyberScan CON 1500



Each mode selection (conductivity, resistivity or total dissolved solids)
has similar setup pages except for TDS mode which includes Setup
P6.0 for setting TDS factor
. Refer to Overview of Setup Pages shown
below.

Mode

Setup Pages Accessible


uS

P1.0 to P15.0 except P6.0

kohm

P1.0 to P15.0 except P6.0

TDS

P1.0 to P 15.0

)

If you do not wish to make a selection or change the option, pressing
std key at any point of time allows you to exit the setup page/mode
and return to the measurement mode.


Overview of Setup Pages in CyberScan CON 1500

P 1.0: View calibration

data

- View the last calibrated

buffer value depending on the
display mode selected.

P 3.0: Set the

Temperature Coefficient

- Select temperature
coefficient value in the range
of 0.000 to 10.000%.

P 5.0: Set single- or

multi-point calibration

- Select ON to activate single-
point calibration, OFF for

multi-point calibration.

P 4.0: Select the

Normalised temperature

- Set the normalised
temperature in the range of
15.0 to 30.0°C.

P 7.0: Set Temperature

Unit

-Set the temperature unit to
be °C or °F.

P 8.0: Set the Baud

Rate

- Set baud rate as

4800, 9600, 19200 or

38400 bps.

P 2.0: Set the cell

constant

- Adjust and select cell
constant to be 1.00, 10.00 or
0.10.

Print

Cell Constant

P 9.0: Set the Parity Bit

- Set parity bit as none(0),

odd(1) or even(2).

P 10.0: Set the Stop Bit

- Set stop bit as one(1) or

two (2).

P 11.0: Select the Print

Data Option

- Select to print current or

stored data.

COEFF

P 6.0: Set TDS Factor

- Set the TDS Factor from a
range of 0.40 to 1.00.

P 14.0: Clear the user

calibration

- Select to clear respective

user calibration in the
current mode.

View

select

Print

Print

Print

clear

clear
BUFFER

STABLE

%

TDS

P 5.0: Select the Auto or

Manual cal options.

(Only applicable for

Conductivity mode)

- Set Auto calibration or

Manual calibration for uS

P 8.0: Set the stability

indicator

- Select ON to activate the

stability indicator, OFF to

de-activate.

OFF

P 13.0: Clear the memory

- Select to clear stored

memory.

P 15.0: Replatinisation

process setting

- Select YES to replatinise

probe, NO to exit. Default

replatinisation duration is 5.00
minutes.