Instruction Manual CyberScan CON 1500

40

3. Press

the

enter key to make the selection and use the ▲/save or

▼/view keys to toggle between NO or YES. Select YES to clear
all the existing buffer values or NO if no change is to be made.

4. Press the enter key to confirm selection OR press std to exit

from this page.

5. Continue to access other setup pages using ▲/save or ▼/view

keys OR press std key to exit from the setup and return to
measurement mode.

)

Depends on the measurement mode you are in, this setup page
clears only the particular mode calibrated value.

Suppose you access the setup pages from TDS measure mode and
you choose to clear the standardised buffer then the meter only
resets the TDS standardisation. Conductivity/ Resistivity
standardisation if any will remain unchanged.

)

When accessing the cell constant, temperature coefficient, TDS
factor, temperature unit selection and stable ON/OFF setup pages,
meter will display the last selected choice.