Thermo Fisher Scientific CyberScan CON 1500 User Manual

Page 58

Advertising
background image

Instruction Manual CyberScan CON 1500

54

Depending on which temperature probe is connected (4-cell/ 2-cell
temperature probe), user can adjust the temperature value using the
▲/save or ▼/view keys. [Window: ± 5°C (9°F)]

If no probe is connected, user can adjust the default temperature
value.

2. Press

▲/save or ▼/view keys to adjust the temperature.

3. Press enter to confirm the reading and return to measurement

mode

OR

4. Press

std key to exit from this setup page.

Note:

ATC: Maximum adjustable temperature offset is ±5.0°C/ 9.0°F.
MTC: Maximum adjustable temperature settings is -5.0°C to 105.0°C

(23.0°F to 221.0°C).