15. METER SPECIFICATIONS
Conductivity
Resistivity
TDS
Range
0.000 to 500.0 mS/cm
0.000 to 99.99 M
Ω
0.000 to 99999 ppm
R1
0.000 to 200.0 uS
0.000 to 20.00K
Ω
0.000 to 200.0 ppm
R2
200.0 uS to 2.000 mS
20.00K to 200.0K
200.0 ppm to 2000 ppm
R3
2.000 mS to 20.00 mS
200.0K to 2.000M
2000 ppm to 20000 ppm
Internal
Range
R4
20.00 mS to 500.0 mS
2.000M to 100.0M
20000 ppm to 99999 ppm
Resolution
0.001 (0.000 to 9.999 uS/cm)
0.01 (10.00 to 99.99 uS/cm)
0.1 (100.0 to 999.9 uS/cm)
0.001 (1.000 to 9.999 mS/cm)
0.01 (10.00 to 99.99 mS/cm)
0.1 (100.0 to 500.0 mS/cm)
0.001 (0.000 to 9.999
Ω)
0.01 (10.00 to 99.99
Ω)
0.1 (100.0 to 999.9
Ω)
0.001 (1.000 to 9.999 k
Ω)
0.1 (100.00 to 999.9 k
Ω)
0.001 (1.000 to 9.999 M
Ω)
0.01 (10.00 to 99.99 M
Ω)
0.001 (0.000 to 9.999 ppm)
0.01 (10.00 to 99.99 ppm)
0.1 (100.0 to 999.9 ppm)
1 (1000 to 99999 ppm)
Accuracy
±0.5% full scale reading
Temperature
range
-5.0 to 105.0°C (23.0 to 221.0°F)
Resolution
0.1°C (0.1°F)
Accuracy
±0.3°C (±0.5°F)
Temperature
Coefficient
0.000 to 10.000
Normalisation
temperature
15.0 to 30.0°C (59.0 to 86.0°F)
TDS Factor
0.40 to 1.00
Cell Constant
0.1, 1.0, 10.0
Display
105 X 75 mm (screen size) custom LCD with contrasted background
Memory
100 data sets
Power
Requirement
Power Adapter 110/ 220VAC, 12V DC, 500mA, centre negative
Inputs and
Outputs
power socket, socket pin (2-cell conductivity), DIN (4-cell conductivity),
2.5 mm phono socket (ATC) and bi-directional RS232
Dimension/
weight
23 X 18 X 6 cm; 950g (without box) / 40X 26 X 9 cm; 1.6kg (boxed)