Instruction Manual CyberScan CON 1500

21

4.2.6 Setup Page 5.0: Set the Single/Multi Point

Calibration mode (SPCAL for all modes)

This setup page allows you to select a single-point or a multi-point
calibration. This option is applicable during all modes of
measurement (Conductivity, Resistivity and Total Dissolved Solids).

1. Access

the

Set the Single/Multi Point Calibration Mode (SPCAL)

setup page by pressing setup during measurement mode
screen.


2. Use the ▲/save or ▼/view keys to scroll until the screen

displays the page as shown below.

)

This page will be displayed only if setup is entered from Resistivity or
TDS measurement mode. However, when setup is entered from the
conductivity measurement mode, this page is not accessible. You
need to enter the Set the Auto Calibration Mode (ACAL) setup page
first and then press the enter key to directly access the Set the
Single/Multi Point Calibration Mode (SPCAL)
setup page as in step 3 .


save

view

press

to set value

press

to

accept

enter