Thermo Fisher Scientific CyberScan CON 1500 User Manual

Page 47

Advertising
background image

Instruction Manual CyberScan CON 1500

43

MIN

4. Use

the

▲/save or ▼/view keys to scroll through the options of

YES and NO for the selection of replatinisation.


5. Press the enter key to confirm selection OR press std to exit

from this page.

6. Continue to access other setup pages using ▲/save or ▼/view

keys OR press std key to exit from the setup and return to
measurement mode.

Replatinisation Timer

1. Select YES to start replatinisation process. “5:00 minutes” will

appear on the main display.















2. Press the enter key to initiate replatinisation. The timer will

countdown from 5.00 minutes on the main display. Upon
completion, the meter will return to the measurement mode
screen.


3. Repeat step 2 if a longer replatinisation is required. A new cell

constant must now be established for the probe. Refer to the
Select the cell constant setup pages.

4. Press

stdby to exit from the replatinisation mode. Meter will be in

standby mode after exit. Press stdby again should you wish to
enter measurement mode.