Thermo Fisher Scientific CyberScan CON 1500 User Manual
Page 47
Instruction Manual CyberScan CON 1500
43
MIN
4. Use
the
▲/save or ▼/view keys to scroll through the options of
YES and NO for the selection of replatinisation.
5. Press the enter key to confirm selection OR press std to exit
from this page.
6. Continue to access other setup pages using ▲/save or ▼/view
keys OR press std key to exit from the setup and return to
measurement mode.
Replatinisation Timer
1. Select YES to start replatinisation process. “5:00 minutes” will
appear on the main display.
2. Press the enter key to initiate replatinisation. The timer will
countdown from 5.00 minutes on the main display. Upon
completion, the meter will return to the measurement mode
screen.
3. Repeat step 2 if a longer replatinisation is required. A new cell
constant must now be established for the probe. Refer to the
Select the cell constant setup pages.
4. Press
stdby to exit from the replatinisation mode. Meter will be in
standby mode after exit. Press stdby again should you wish to
enter measurement mode.
save
view