Thermo Fisher Scientific CyberScan CON 1500 User Manual

Page 16

Advertising
background image

Instruction Manual CyberScan CON 1500

12

press

to

accept

enter

S

STD Solution

u

4. When you have scrolled through all the cell constants, you will

automatically come to the screen shown below. This screen
displays the meter’s calibrated value.


5. Press enter key to go to the next setup page P2.0. Press std

key if you wish to return to measurement mode.

6. Continue to access other setup pages using ▲/save or ▼/view

keys OR press std key to exit from the setup mode and return to
measurement mode.

































)

If you enter the setup mode from Conductivity measurement mode,
calibration data will be in µS or mS. Similarly, if you enter the setup
mode from TDS measurement mode, calibration data will be in ppm
or ppt.