Thermo Fisher Scientific CyberScan CON 1500 User Manual
Page 51
![background image](/manuals/370649/51/background.png)
Instruction Manual CyberScan CON 1500
47
5.2.
Conductivity standardisation (Auto)
Refer to Section 4.2.5: Set the Auto Calibration Mode (ACAL) on
Page 19 to enable the automatic calibration of conductivity feature in
this meter.
Standard buffer values: 84.0 uS/ 1.413 mS/ 12.88 mS/ 111.8 mS
1. Immerse the electrode into the standard solution you wish to
calibrate.
2. Press the std key. The lower display will show the standard
buffer value closest to the measured value as shown in the next
page.
3. Press enter to initiate the standardisation and on successful
completion meter returns to measurement mode
OR
Press std to exit from the standardisation page.
)
Standardisation window provided is ±40% of current reading. If the
standard value set is not within the window then meter displays
“Electrode Error” message after initiate the standardisation.
ELECTRODE ERROR
Check Electrode/ Buffers/ Temperature
Press enter to return to measurement mode. In this case meter will
not accept the standardisation,
If the standardisation is successful, then meter displays buffer icon in
the middle part of the display in the measurement mode.