Chemical Injection Applications

For water line applications with a chemical injection point, install the meter upstream of the injection point This eliminates

any meter performance issues

Figure 10: Chemical Injection Point Upstream of Meter

If a meter must be installed downstream of a chemical injection connection, the distance between the meter and the

injection point should be between 50 and 100 feet (15 and 30 meters) The distance must be long enough to allow the water/

chemical solution to reach the meter in a complete, homogeneous mixture If the injection point is too close, the meter senses

the two different conductivities for each liquid This will likely result in inaccurate measurements The injection method –

spaced bursts, continuous stream of drips or liquid or gas – can also affect downstream readings by the meter

Figure 11: Chemical Injection Point Downstream of Meter

Sometimes, due to circumstances, it’s difficult to specify the exact downstream placement distances Contact Badger Meter

Technical Support to review your application if necessary