Menu Manager Configuration Errors

Error

Description

Recommended Action

140

Output 3: Configuration Error – Full

scale frequency exceeds limits of

relay (1000 Hz)

Reduce full scale frequency setting of output when hardware is configured for relay operation

150

Output 3: Configuration Error – Full

scale frequency exceeds limits (10

kHz)

Reduce full scale frequency setting of output when hardware is configured for open collector

operation

170

Output 1/2: Output Type
Configuration Error

This error is observed when the function type is 24V DC and the output type is changed from
Normally Open to Normally Closed It is required for 24V DC output operation that the output type be
Normally Open

171

Output 1/2: Output Type
Configuration Error

This error is observed when the function type is ADE and the output type is changed from Normally
Open to Normally Closed It is required for ADE operation that the output type be Normally Open

190

Full Scale Flow: Entered Value
exceeds limits

Value entered exceeds the absolute maximum flow the meter supports Reduce the value for this
parameter or consider increasing pipe diameter

Display Error / Status Messages

Error Message

Possible Cause

Recommended Action

Err: Detector

No detector connection with amplifier

Check detector and cable connections in accordance with this manual

Connection between amplifier and detector

Contact Technical Support

Supply voltage too low

Contact Technical Support

Grounded coils in meter

Contact Technical Support

Water in detector

Contact Technical Support

Err: Empty pipe

Pipe may not be full

Make sure all trapped air is out of system

If fluid or fluid conductivity recalibrate the parameter

Err: Full scale

Actual flow rate is exceeding programmed flow Reduce flow rate or increase the programmed full scale value by more than 5%

Err: AD-Range

AD-Converter is exceeding signal limits

Check the grounding scheme of the meter installation See grounding section in this

manual Verify pipe is not empty

Err: AD-INT

Initialization of AD-Converter unsuccessful

Contact Technical Support

Err: Rollover

Rollover counters have exceeded limit

Clear all totalizers

Err: Rollover Status

Totalizer rollover has occurred

Reload totalizer then clear all totalizers

Err: Simulation

I/O simulator is enabled

Disable simulator in I/O menu

M-Series® M2000 Electromagnetic Flow Meter