Clear Totals

T1

S

The uni-directional totalizer is reset within the menu manager or through remote communications

Clearing T1 also clears the associated rollover counter

T2

U

The uni-directional totalizer is reset within the menu manager, through remote communications or with

properly-configured digital input (function = remote reset) Clearing T2 also clears the associated rollover

counter

T+

S

The bi-directional forward flow totalizer is reset within the menu manager or through remote

communications Clearing T+ also clears the associated rollover counter

T

S

The bi-directional reverse flow totalizer is reset within the menu manager or through remote

communications Clearing T– also clears the associated rollover counter

Tn

S

The bi-directional net totalizer, when reset, clears both the forward and the reverse flow totalizers (T+ and

T–) This is reset within the menu manager or through remote communications Clearing Tn also clears the

associated rollover counter

PS

S

The batch totalizer is reset to the configured preset amount value It is reset within the menu manager,

remote communications or through a properly-configured digital input (function = batch reset)

Tpwroff

S

The totalizer accumulating meter time without external power is reset with the menu manager or through

remote communications

Communication: Port Settings

Interface
[MODBUS RTU]

S

The Interface parameter lets you configure how the RS232 communication port will be used

• MODBUS RTU

• Remote menu (RDI – Remote Display Interface)

• Primo 3 x

• Disable port

The remote menu interface will check for display updates once a second If a change is detected, the

display contents will be transmitted in ASCII format over the RS232 communication port The remote

menu interface also allows for menu navigation and control of the meter as if using the external push

buttons Keyboard control characters such as <UP>,<DWN>, and <ENTER> are supported to navigate the

menus
The Primo 3 x interface will emulate the legacy Primo 3 x Protocol This protocol will transmit an ASCII

string in the following format every 500 ms:

“RATE;0 0000; GPM; TOT1;150 0000; USG ; TOT2;150 0000; USG ;” – For Unidirectional Mode

“RATE;0 0000; GPM; TOT+;10 0000; USG ; TOT-;50 0000; USG ;” – For Bidirectional Mode
To change the Interface follow these steps from the Port A Settings menu:

1 Select Interface to view the Interface display
2 Select the desired interface
3 Press [E] to save and return to the Port A Settings menu