Inputs/Outputs

Digital Output Preset Amount

[0 0]

S

Preset amount lets you set the reset value for the associated PS totalizer when the

digital input is set to Batch Reset
To change the preset amount, follow these steps from the Inputs/Outputs menu:

1 Select Digital Output 1, 2, 3 or 4 and press [E] to open the Digital Output

menu

2 From the Digital Output menu select Preset Amount, and press [E] to open the

Preset Amount display

3 Enter the preset amount value Press [E] to save the new parameter and return

to the Digital Output menu

NOTE:

N

You can only set one Preset Amount If you set the Preset Amount for Digital

Output 1, it will be the same for 2, 3 and 4

Set Point Min.
[0%]

S

This parameter establishes, as a percentage of full scale flow, the threshold at which

the output alarm will be activated Flow rates below the threshold will activate the

output alarm
To change the set point minimum, follow these steps from the Inputs/Outputs menu:

1 Select Digital Output 1, 2, 3 or 4 and press [E] to open the Digital Output

menu

2 From the Digital Output menu select Set Point Min., and press [E] to open the

Set Point Min. display

3 Enter the set point minimum value Press [E] to save the new parameter and

return to the Digital Output menu

Set Point Max.
[100%]

S

This parameter establishes, as a percentage of full scale flow, the threshold at which

the output alarm will be activated Flow rates above the threshold will activate the

output alarm
To change the maximum set point, follow these steps from the Inputs/Outputs menu:

1 Select Digital Output 1, 2, 3 or 4 and press [E] to open the Digital Output

menu

2 From the Digital Output menu select Set Point Max. and press [E] to open the

Set Point Max. display

3 Enter the set point maximum value and press [E] to save the new parameter

and return to the Digital Output menu

M-Series® M2000 Electromagnetic Flow Meter