Meter Orientation

Mag meters can operate accurately in any pipeline orientation and can measure volumetric flow in forward and reverse

directions

NOTE:

N

A "Forward Flow" direction arrow is printed on the detector label

Vertical Placement

Mag meters perform best when placed vertically, with liquid flowing upward and meter electrodes in a closed, full pipe

Figure 6: Vertical Placement

Vertical placement allows the pipe to remain completely full, even in low flow, low pressure applications, and it prevents

solids build-up, sediment deposit and accumulation on the liner and electrodes

NOTE:

N

Carefully observe the “Forward Flow” label on the meter body and install the meter accordingly When

installed vertically, rotate amplifier so that cable glands are facing down

Horizontal Placement

M2000 meters are equipped with an Empty Pipe Detection feature If an electrode mounted in the pipe is not covered by fluid

for five seconds, the meter will display an Empty Pipe Detection condition The meter will send out an error message and

stop measuring flow When the electrode is again covered with fluid, the error message disappears and the meter will begin

measuring
When installing the meter on a horizontal pipe, mount the detector to the pipe with the flow-measuring electrode axis in a

horizontal plane (three and nine o’clock) This placement helps prevent solids build-up, sediment deposit and accumulation

on the electrodes

Electrode

Plane

RIGHT

Electrode

Plane

WRONG

Figure 7: Horizontal Placement

Installation & Operation Manual

Page 9