Measurements

Totalizer Unit

[USG]

U

The Totalizer Unit parameter establishes the units of measure for the totalizers
To change the Totalizer Unit value, follow these steps from the Measurements menu:

1 Select Totalizer Unit to view the Totalizer Unit display
2 Position the arrow next to one of the following totalizer units:

Code

Totalizer Unit

Code

Totalizer Unit

L

Liters

UKG

Imperial Gallons

HL

HectoLiters

Lb

Pounds

M

3

Cubic Meters

Oz

Fluid Ounces

CFt

Cubic Feet

Aft

Acre Feet

USG

U S Gallons

BBL

Barrel

MG

MegaGallons

3 Press [E] to save the totalizer unit and return to the Measurements menu

Full Scale Flow

[Factory Set]

U

The Full Scale Flow parameter sets the maximum flow the system is expected to measure This

parameter has influence on other system parameters These parameters include:

• Frequency Output – Full scale frequency is observed at full scale flow
• Low Flow Cutoff – Changes to full scale flow affect the measuring cut-off threshold of the meter
• Alarm Outputs – Changes to full scale flow adjusts the thresholds for generating set point alarms
• Pulse Outputs – Changes to full scale flow adjusts the pulse frequency and duty cycle
• Analog Outputs – Changes to full scale flow adjusts the interpretation of the analog output signal

Change the full scale flow based on the meter size and the application’s requirements Verify that the

full scale flow falls within the meter’s suggested flow range limits In terms of flow velocity, the meter's

limits are from 0 1 to 39 4 feet/second
The full scale flow is valid for both flow directions

NOTE:

N

If the flow rate exceeds the full scale setting, an error message indicates that the configured full

scale range has been exceeded However, the meter will continue to measure This will affect

the latency of the pulse outputs and possibly cause overflow Furthermore, the analog output

may also be placed in alarm mode

To change the Full Scale Flow, follow these steps from the Measurements menu:

1 Select Full Scale Flow to view the Full Scale Flow display
2 Set the full scale flow value to the desired setting
3 Press [E] to save the full scale flow value and return to the Measurements menu

M-Series® M2000 Electromagnetic Flow Meter