Wiring for remote configuration, Empty pipe detection considerations – Badger Meter M-Series M2000 User Manual

Page 17

Advertising
background image

Wiring for Remote Configuration

44

40

44

46

44

45

13

12

11

shield (black)
pink

brown
shield (black)
white

shield (black)
yellow
green

green
yellow
shield (black)

white
shield (black)

brown

pink
shield (black)

E1
ES
E2
RS
EP
ES

CS C2 C1

Figure 17: Wiring for Remote Configuration

Remote style M2000 amplifier models can be ordered with standard cables measuring 15, 30, 50 and 100 feet In addition,

cables up to 500 feet are available

Junction Box

Amplifier

Connection No.

Description

Wire Color

Connection

11

Coil

Green

C1

12

Coil

Yellow

C2

13

Main Shield

Black (Red Ferrule)

CS

45

Electrode

White

E1

44*

Electrode Shield

Black

ES

46

Electrode

Brown

E2

40

Empty Pipe

Pink

EP

44*

Empty Pipe Shield

Black

ES

*Connections with the No 44 are lying on the same potential

Empty Pipe Detection Considerations

Take into account the following cable length and conductivity requirements, if you will be using empty pipe detection

Cable Length (Feet)

Minimum Conductivity Required (µS/cm)

0*

5

100

20

500

100

* Meter Mount

Installation & Operation Manual

Page 17