Quick Setup

Full Scale Flow

Full Scale Flow sets the maximum flow the system is expected to measure This parameter has

influence on other system parameters These parameters include:

• Frequency Output – Full scale frequency is observed at full scale flow
• Low Flow Cutoff – Changes to full scale flow affect the measuring cut-off threshold of the meter
• Alarm Outputs – Changes to full scale flow adjusts the thresholds for generating set point alarms
• Pulse Outputs – Changes to full scale flow adjusts the pulse frequency and duty cycle
• Analog Outputs – Changes to full scale flow adjusts the interpretation of the analog output signal

Change the full scale flow based on the meter size and the application’s requirements Verify that the

full scale flow falls within the meter’s suggested flow range limits In terms of flow velocity, the meter's

limits are from 0 1 to 39 4 feet/second
The full scale flow is valid for both flow directions

NOTE:

N

If the flow rate exceeds the full scale setting, an error message indicates that the configured

full scale range has been exceeded However, the meter will continue to measure This will

affect the latency of the pulse outputs and possibly cause overflow Furthermore, the analog

output may also be placed in alarm mode

To set or change the Full Scale Flow, follow these steps from the Quick Setup menu:

1 Select Full Scale Flow to view the Full Scale Flow display
2 Set the full scale flow value to the desired setting
3 Press [E] to save the full scale flow value and return to the Measurements menu

Low Flow Cutoff

Low Flow Cutoff defines the threshold at which flow measurement will be forced to zero The cutoff

value can be set from 0% to 10% of the full scale flow Increasing this threshold will help prevent false

readings during “no flow” conditions possibly caused by pipe vibration or inherent system noise
To change Low Flow Cutoff, follow these steps from the Low Flow Cutoff display

1 Set the Low Flow Cutoff value to the desired setting, between 0% and 10%
2 Press [E] to save the value