Badger Meter M-Series M2000 User Manual

Page 44

Advertising
background image

ADE

Protocol Type

The protocol type enables the ADE interface

Selecting S1 or S2 will automatically configure

the Digital Input and Digital Output #1 for ADE

operation Manually configuring the input and

output for ADE operation is not allowed and will

result in a on-screen error
See "ADE Interface" on page 52 for further

details
S1 – Standard ADE protocol
S2 – Enhanced ADE protocol, provides additional

digital information
Disabled – disables and removed ADE

configuration

Dial Type

4 – 8 dials

Analog Calibrate

Custom Settings
[Zero Scale: 0 mA]

[Full Scale: 0 mA]

S

To set the analog calibration custom settings, follow these steps from the

Advanced menu:

1 Select Analog Calibrate to view the Analog Calibrate menu
2 Select Custom Settings to view the Custom Settings display
3 Select one of the following:

• Offset 4 mA
• Offset 20 mA

4 Configure desired offset
5 Press [E] to save the option and to return to the Custom Settings menu
6 Press [E] to return to the Analog Calibrate menu

Factory Settings
[Factory Set]

S

To change the analog calibration factory settings, follow these steps from the

Advanced menu:

1 Select Analog Calibrate to view the Analog Calibrate menu
2 Select Factory Settings to view the Factory Settings display
3 Select one of the following:

• Calibration Point A
• Calibration Point B

4 Set the calibration point to the measured output current
5 Press [E] to save the option and to return to the Factory Settings menu
6 Press [E] to return to the Analog Calibrate menu

M-Series® M2000 Electromagnetic Flow Meter