Chapter 2: prerequisites – Labconco Protector ClassMate Laboratory Hoods 69716 User Manual

Page 13

Advertising
background image

9

Product Service 1-800-522-7658

Chapter 2: Prerequisites

Total Exhaust at 18" Sash Opening

60% open

Catalog

Number

Standard Model Description

Face

Velocity

Exhaust

Volume

Static

Pressure

Loss

69704-00

48" Hood without fixtures & duplexes

60 fpm

340 CFM

0.09" H

2

O

80 fpm

460 CFM

0.16" H

2

O

100 fpm

570 CFM

0.24" H

2

O

69704-01

48" Hood with 4 fixtures & 2 duplexes

60 fpm

340 CFM

0.09" H

2

O

80 fpm

460 CFM

0.16" H

2

O

100 fpm

570 CFM

0.24" H

2

O

69705-00

60" Hood without fixtures & duplexes

60 fpm

440 CFM

0.14" H

2

O

80 fpm

590 CFM

0.26"H

2

O

100 fpm

730 CFM

0.39" H

2

O

69705-01

60" Hood with 4 fixtures & 2 duplexes

60 fpm

440 CFM

0.14" H

2

O

80 fpm

590 CFM

0.26"H

2

O

100 fpm

730 CFM

0.39" H

2

O

69706-00

72" Hood without fixtures & duplexes

60 fpm

540 CFM

0.19" H

2

O

80 fpm

710 CFM

0.33" H

2

O

100 fpm

890 CFM

0.51" H

2

O

69706-01

72" Hood with 4 fixtures & 2 duplexes

60 fpm

540 CFM

0.19" H

2

O

80 fpm

710 CFM

0.33" H

2

O

100 fpm

890 CFM

0.51" H

2

O

Total Exhaust at 50% Open Sash

Horizontally or Vertically

Catalog

Number

Reduced Air Volume Model Description

Face

Velocity

Exhaust

Volume

Static

Pressure

Loss

69714-00

48" Single Sash Hood without monitor, fixtures & duplexes

60 fpm

280 CFM

0.06" H

2

O

80 fpm

380 CFM

0.11"H

2

O

100 fpm

470 CFM

0.16" H

2

O

69714-01

48" Single Sash Hood with digital airflow monitor, 4 fixtures & 2
duplexes

60 fpm

280 CFM

0.06" H

2

O

80 fpm

380 CFM

0.11"H

2

O

100 fpm

470 CFM

0.16" H

2

O

69715-00

60" A-Style Sash Hood without monitor, fixtures & duplexes

60 fpm

330 CFM

0.08" H

2

O

80 fpm

440 CFM

0.14" H

2

O

100 fpm

550 CFM

0.22" H

2

O

69715-01

60" A-Style Sash Hood with digital airflow monitor, 4 fixtures & 2
duplexes

60 fpm

330 CFM

0.08" H

2

O

80 fpm

440 CFM

0.14" H

2

O

100 fpm

550 CFM

0.22" H

2

O

69716-00

72" A-Style Sash Hood without monitor, fixtures & duplexes

60 fpm

410 CFM

0.11" H

2

O

80 fpm

550 CFM

0.20" H

2

O

100 fpm

680 CFM

0.30" H

2

O

69716-01

72" A-Style Sash Hood with digital airflow monitor, 4 fixtures & 2
duplexes

60 fpm

410 CFM

0.11" H

2

O

80 fpm

550 CFM

0.20" H

2

O

100 fpm

680 CFM

0.30" H

2

O

Proper blower selection can be determined from these
exhaust requirements and the total system static
pressure loss. Consult Labconco Sales Engineering
Team if you need help sizing a blower.

Advertising