Chapter 2: Prerequisites
Total Exhaust at 18" Sash Opening
60% open
Catalog
Number
Standard Model Description
Face
Velocity
Exhaust
Volume
Static
Pressure
Loss
69704-00
48" Hood without fixtures & duplexes
60 fpm
340 CFM
0.09" H
2
O
80 fpm
460 CFM
0.16" H
2
O
100 fpm
570 CFM
0.24" H
2
O
69704-01
48" Hood with 4 fixtures & 2 duplexes
60 fpm
340 CFM
0.09" H
2
O
80 fpm
460 CFM
0.16" H
2
O
100 fpm
570 CFM
0.24" H
2
O
69705-00
60" Hood without fixtures & duplexes
60 fpm
440 CFM
0.14" H
2
O
80 fpm
590 CFM
0.26"H
2
O
100 fpm
730 CFM
0.39" H
2
O
69705-01
60" Hood with 4 fixtures & 2 duplexes
60 fpm
440 CFM
0.14" H
2
O
80 fpm
590 CFM
0.26"H
2
O
100 fpm
730 CFM
0.39" H
2
O
69706-00
72" Hood without fixtures & duplexes
60 fpm
540 CFM
0.19" H
2
O
80 fpm
710 CFM
0.33" H
2
O
100 fpm
890 CFM
0.51" H
2
O
69706-01
72" Hood with 4 fixtures & 2 duplexes
60 fpm
540 CFM
0.19" H
2
O
80 fpm
710 CFM
0.33" H
2
O
100 fpm
890 CFM
0.51" H
2
O
Total Exhaust at 50% Open Sash
Horizontally or Vertically
Catalog
Number
Reduced Air Volume Model Description
Face
Velocity
Exhaust
Volume
Static
Pressure
Loss
69714-00
48" Single Sash Hood without monitor, fixtures & duplexes
60 fpm
280 CFM
0.06" H
2
O
80 fpm
380 CFM
0.11"H
2
O
100 fpm
470 CFM
0.16" H
2
O
69714-01
48" Single Sash Hood with digital airflow monitor, 4 fixtures & 2
duplexes
60 fpm
280 CFM
0.06" H
2
O
80 fpm
380 CFM
0.11"H
2
O
100 fpm
470 CFM
0.16" H
2
O
69715-00
60" A-Style Sash Hood without monitor, fixtures & duplexes
60 fpm
330 CFM
0.08" H
2
O
80 fpm
440 CFM
0.14" H
2
O
100 fpm
550 CFM
0.22" H
2
O
69715-01
60" A-Style Sash Hood with digital airflow monitor, 4 fixtures & 2
duplexes
60 fpm
330 CFM
0.08" H
2
O
80 fpm
440 CFM
0.14" H
2
O
100 fpm
550 CFM
0.22" H
2
O
69716-00
72" A-Style Sash Hood without monitor, fixtures & duplexes
60 fpm
410 CFM
0.11" H
2
O
80 fpm
550 CFM
0.20" H
2
O
100 fpm
680 CFM
0.30" H
2
O
69716-01
72" A-Style Sash Hood with digital airflow monitor, 4 fixtures & 2
duplexes
60 fpm
410 CFM
0.11" H
2
O
80 fpm
550 CFM
0.20" H
2
O
100 fpm
680 CFM
0.30" H
2
O
Proper blower selection can be determined from these
exhaust requirements and the total system static
pressure loss. Consult Labconco Sales Engineering
Team if you need help sizing a blower.