Labconco Protector ClassMate Laboratory Hoods 69716 User Manual

Page 25

Advertising
background image

21

Product Service 1-800-522-7658

Chapter 3: Getting Started


The identification plate, model number, serial number,
and electrical connection boxes are accessible from the
front of the fume hood by removing the front panel.

The remote blower style Protector ClassMate Hood is
fully wired internally for 115 Volt, 60 Hz, 15 Amp
electrical service. The number of circuits varies
depending on the model. The Protector ClassMate
Hood models do not feature internal wiring or electrical
components excluding the fluorescent light assembly.
All of the hood’s electrical connections are terminated
at the internal boxes for hook-up by a qualified
electrician. The specific internal boxes for hook-up are
the lights, blower, and 115V, 60 Hz, 15A duplex
outlets. Refer to the wiring diagram for your Protector
ClassMate in Appendix C: Protector ClassMate Fume
Hood Specifications.




All wiring for the enclosure MUST be
performed by a licensed electrician
and conform to all local codes. In
most cases, the hood will require the
use of shielded conduit to protect the
wiring into the hood. The grounding
connection shall not be made to the
terminal box cover.

Figure 3-7