Product Service 1-800-522-7658
Chapter 3: Getting Started
Your Protector ClassMate Fume Hood has been
certified at the factory per ASHRAE 110-1995. In this
procedure Labconco tests the “as manufactured” hood
for average face velocity readings, smoke visualization,
and tracer gas containment readings less than 0.1 part
per million (PPM) with a mannequin. For “field use”
ASHRAE testing contact Labconco Sales Engineering
Team for a certified on-site contractor.
NOTE: Face velocity profiles and
smoke testing should be done
periodically to ensure proper safe
performance.
Your Next Step
After your Fume Hood has been installed and certified,
you are ready to proceed to Chapter 4: Theory of
Operation and Safety Precautions.
!