Chapter 5: Using Your Protector ClassMate Fume Hood

Working in your Protector
ClassMate Fume Hood

Planning

• Thoroughly understand procedures and

equipment required before beginning work.

• Arrange for minimal disruptions, such as room

traffic or entry into the room while the hood is
in use.


Start-up

• Turn on fluorescent light and hood blower.

• Slowly raise the sash.

• Check the baffle air slots for obstructions.

• Allow the hood to operate unobstructed for 5

minutes.

• Wear a long sleeved lab coat with knit cuffs and

rubber gloves. Use protective eyewear. Wear a
protective mask if appropriate.


Wipe-Down

• Raise the sash to its full open position.

• If necessary, wipe down the interior surfaces of

the hood with dish soap and glass cleaner and
allow to dry.


Loading Materials and Equipment

• Only load the materials required for the

procedure. Do not overload the hood.

• Do not obstruct the front airfoil, or rear baffle

slots.

• Large objects should not be placed close

together.

• After loading the hood, wait one minute to

purge airborne contaminants from the work
area.


Work Techniques

• Keep all materials at least 6 inches inside of the

sash, and perform all contaminated operations
as far to the rear of the work area as possible.

• Segregate all clean and contaminated materials

in the work area.