Model Number 6970400

6970401

6970500
6970501

6970600
6970601

6971400
6971401

6971500
6971501

6971600
6971601

Model Description

4'

Standard

5'

Standard

6'

Standard

4' w/ Sash

Stop Energy

Saving

5' A-Style

Energy

Saving

6' A-Style

Energy

Saving

Sash Type Vertical-

Rising

Pivoting

Vertical-

Rising

Pivoting

Vertical-

Rising

Pivoting

Vertical-

Rising

w/Sash Stop

A-Style

Combination

A-Style

Combination

Operating Sash Height (in)

18.00

18.00

18.00

15.00 30.00 30.00

Operating Sash Width (in)

45.00

57.00

69.00

45.00 26.00 32.00

Total Open Area (Sq. Ft.) Includes

bypass air under airfoil

5.70

7.30

8.90

4.70 5.50 6.80

Average Face Velocity (FPM)

100±10

100±10

100±10

100±10

100±10

100±10

Total Air Volume Displacement (CFM)

570

730

890

470

550

680

Hood Static Pressure Loss (in H

2

0)

0.24

0.39

0.51

0.16 0.22 0.30

Average Face Velocity (FPM)

80±10

80±10

80±10

80±10

80±10

80±10

Total Air Volume Displacement (CFM)

460

590

710

380

440

550

Hood Static Pressure Loss (in H

2

0)

0.16

0.26

0.33

0.11 0.14 0.20

Average Face Velocity (FPM)

60±10

60±10

60±10

60±10

60±10

60±10

Total Air Volume Displacement (CFM)

340

440

540

280

330

410

Hood Static Pressure Loss (in H

2

0)

0.09

0.14

0.19

0.06 0.08 0.11


The following calculation is used:
Total Air Volume Displacement (CFM)=Total Open
Area (Sq. Ft.) *Average Face Velocity (FPM).