Chapter 2: Prerequisites

Electrical Requirements

The Protector ClassMate Hood models feature internal
wiring for the fluorescent light assembly and light
switch. All hood models with 115V, 60 Hz duplex
outlets are terminated at the internal boxes for hook-up
by a qualified electrician. The blower switch, and light
switch wires are also terminated at the internal boxes
for hook-up by a qualified electrician. Refer to Chapter
3: Getting Started
and Appendix C: Protector
ClassMate Specifications
for the wiring diagram for
proper electrical installation. All models have 115
VAC, 60 Hz, 12 Amp requirements.

Service Line Requirements

All service lines to the laboratory fume hood should be
¼ inch outside diameter, copper (brass for gas), and
equipped with an easily accessible shut-off valve,
should disconnection be required. If the service line
pressure exceeds 40 PSI, it must be equipped with a
pressure regulator to reduce the line pressure.

On hood models with service fixtures, three of the
fixtures are pre-plumbed using 1/4" copper tubing
between both the valve, and the hose connector. The
upper right hand position fixture is pre-plumbed using
1/4" brass tubing between both the valve, and the hose
connector for usage with natural gas. The supply lines
shall be installed from under the hood and terminated at
the valves by a qualified contractor.

Space Requirements

The dimensions for the different models are shown in
Appendix B: Protector ClassMate Dimensions.

Your Next Step

After you have determined that the location you have
selected accommodates the installation and operational
requirements of your fume hood, you are ready to begin
installation. Proceed to Chapter 3: Getting Started.