Chapter 7: Modifying Your Protector ClassMate Fume Hood



Step 1: Remove the gray blank wall plate from the hood

corner. It is attached to the hood with silicone
sealant. You will probably damage the blank
wall plate upon removal.


Step 2: Remove the 2 x 4 handy box cover on the top of

the fume hood from the nearest junction box to
the front.


Step 3: Install the cable connectors on the junction box,

one for inlet power, and one for outlet power.


Step 4: Route power from the junction box to the front

corner post using the outlet harness. Route
cable through the slot in the corner post.


Step 5: Wire outlet cable to 115V duplex receptacle and

attach receptacle to front corner post with two
#6-32 supplied screws.


Step 6: Attach gray duplex receptacle wall plate to the

duplex using the #6-32 black oval head screw.


Step 7: Wire 115V, 15A supply line to the outlet

harness. Be sure the line is properly grounded
to the junction box.


Step 8: Circuit test 115V electrical duplex outlet.

Step 9: Replace the 2 x 4 handy box cover when testing

is complete.


Your Next Step


After you understand the modifying procedures, you are
ready to proceed to Chapter 8: Troubleshooting.