Chapter 7: Modifying Your Protector ClassMate Fume Hood

Installing Exhaust Adapter
for Back to Back Hoods

The common exhaust adapter P/N 6987000 allows you to
duct two back-to-back Protector ClassMate Fume Hoods
from two 8" (8.62" O.D. pipe) exhaust stacks to one 12"
(12.00" O.D. pipe) exhaust stack. (See Figure 7-3)

Step 1. Simply place the adapter on both hood exhausts

and seal with silicone sealant.


Step 2. Attach the 12" (12.00" O.D.) pipe to your

blower exhaust.


Installing Guardian Digital
Airflow Monitor

The Digital Airflow Monitor P/N 97432-03 allows you to
continuously monitor face velocity through the fume hood
opening. (See Figure 7-4, Figure 7-5, and Figure 7-6)