Labconco Protector ClassMate Laboratory Hoods 69716 User Manual
Page 20
![background image](/manuals/738680/20/background.png)
16
Product Service 1-800-522-7658
Chapter 3: Getting Started
Follow the instructions as shown in Figures 3-1, 3-2,
and 3-3
Figure 3-1
Figure 3-2
Figure 3-3
1. Remove and discard the
cardboard corner posts
(Figure 3-1).
2. Remove strap that secures
sash. For A-Style
combination sashes
disassemble from skid and
attach to hood per Step 7.
3. Remove the access panels and after
removal, save the access panels and
hardware for re-attachment. (Figure
3-2)
4. Remove and discard the screws and
washers that are located behind the
access panel that attaches the hood
to the skid. Not Shown.
5. Re-attach access panels from Step 3.
(Figure 3-2).
6. Ensure that the cable
on each side is
properly seated in
the front and rear
pulleys. (Figure 3-
3). Releasing the
sash weights with
the cables
improperly
positioned can cause
damage.
7. IMPORTANT: A-
Style Combination
Sash version install
horizontal glass
sashes before Step
8.
8. Release the load
on the bolt by
lifting up slightly
on the cable.
Remove the bolts
straight out of the
rear posts. 2 bolts
per each side to
be removed.
(Figure 3-3)
9. Sash shall raise
smoothly.
!