Chapter 7: Modifying Your Protector ClassMate Fume Hood


2. Mount the WaterSaver valve, listed below, through

the 1.25" diameter hole and tighten the supplied
round locking ring.


Part Number

Description

69837-00

WaterSaver rigid gooseneck and valve

69837-01

WaterSaver (VAC) Connector and valve

69837-02

WaterSaver (AIR) Connector and valve

69837-03 WaterSaver

(GAS)

Connector and valve

69837-04

WaterSaver Swivel Gooseneck and valve


3. Mount two each 3/8" pipe x 1/4" tube male

connectors P/N 1410701 to the WaterSaver valve
with pipe sealant P/N 1577900 and tighten the male
connectors approximately one turn past hand tight.

4. Mount the WaterSaver outlet connector or

gooseneck on the sidewall through the 3/4"
diameter hole and tighten the supplied nut and
washer.

5. Mount the 3/8" female x 3/8” male street elbow P/N

1402400 to the outlet fitting. Apply pipe sealant
P/N 1577900 to the threads. Tighten the fitting with
a wrench or pliers.

6. Mount the 3/8" pipe x 1/4" tube female connector

P/N 4890300 to the street elbow. Apply pipe
sealant P/N 1577900 to the threads. Tighten the
fitting with a wrench or pliers.

7. Connect one end of the outlet service tube, part

numbers listed below, to the female connector
fitting, from Step 5, by pressing the tube all the way
into the fitting, and through the two ferrules, turning
the tubing nut hand tight, and then approximately
one more turn using a 9/16" wrench.

8. Connect the other end of the outlet service tube, part

number listed below, to the WaterSaver valve male
connector fitting by pressing the tube all the way
into the fitting and through the two ferrules, turning
the tubing nut hand tight, and then approximately
one more turn using 9/16” wrench. Note the flow
direction on the valve.



Advertising