Product Service 1-800-522-7658

Chapter 3: Getting Started

The exhaust connection on your hood
has been designed for 8" nominal pipe
(8.62" O.D.) to allow for minimum static
pressure loss with proper transport
velocities away from the hood. It is
acceptable to combine the air from two
back to back hoods with a 12" x 8" x 8"
dual exhaust adapter that would
basically have two 8" nominal exhaust
inlets and one 12" (12.00" O.D.)
nominal exhaust outlet. See Appendix
D: Protector ClassMate Accessories
for
dual exhaust adapter. Refer to exhaust
requirements in Chapter 3 Fume Hood
Components
on exhaust volumes for
your model to properly size the blower.
Consult Labconco Sales Engineering
Team should you require help sizing
your blower for the exhaust volume and
total system static pressure loss.

Connecting the Electrical
Supply Source to the
Protector ClassMate Fume
Hood


Prior to connecting any electrical wiring to the fume
hood structure, refer to the hood identification plate for
the proper electrical characteristics of your specific
model.












WARNING: The building electrical supply
system for Protector ClassMate Hoods should
include overload protection. A switch or
circuit breaker should be in close proximity to
the equipment and within easy reach of the
operator. The switch or circuit breaker is to
be marked as the disconnecting device for the
equipment.

!