Refer to the following table if your fume hood fails to
operate properly. If the suggested corrective actions do
not solve your problem, contact Labconco for
additional assistance.

PROBLEM

CAUSE

CORRECTIVE ACTION

Remote blower and
lights won’t turn on

Wires not connected
at junction boxes or
switches.

Check connection of switches.

Check connection to control box on top
of unit.

Circuit

breakers

tripped in building
electrical supply.

Reset circuit breakers.

Remote Blower
won’t turn on but
lights work

Blower wiring is
disconnected.

Inspect blower wiring, and switch.

Blower

motor

is

defective.

Replace blower motor.

Fume Hood blower
turns on but lights
don’t work

Lamp not installed
correctly.

Inspect lamp installation.

Lamp is defective.

Replace lamp.

Lamp

circuit

breaker

in building is tripped.

Reset the lamp circuit breaker.