Operation

Persistent Switch Control : Exiting Persistent Mode

5. The persistent switch heater is heated for the preset heating time

as set by the PSwitch Heated Time variable

1

.

Note

The magnet voltage (Vm) is monitored during switch heating. If the

voltage is greater than, 0.5 V, the Model 430 Programmer will beep

and display a message to indicate a mismatch between the magnet

current and power supply current:

This mismatch in current indicates the magnet current (and

therefore the field) has decayed significantly during the time

the magnet was in persistent mode. Since this is not a critical

error

2

, after 15 seconds the screen will change back to the

default display:

6. After the Model 430 Programmer has completed this persistent

switch heating operation, the display reads:

1. Refer to section 3.10.2.9 on page 55. The (4) in the display indicates the number of

seconds remaining in the heating cycle (4 in this example).

2. This current mismatch could be indicative of a problem with the magnet persistent

joints.

+50.00 A

Mode: Heating Switch (4)

+3.50 Vs

PSwitch Heater: ON

+50.00 A

Mode: Heating Switch (4)

+3.50 Vs

*** Current Mismatch ***

+50.00 A

P

Mode: Paused

+3.50 Vs

PSwitch Heater: ON

+50.00 A

P

Mode: Magnet in Driven Mode

+3.50 Vs

Press ENTER to continue