Shunt Terminals

The connectors route the incoming signals from the Magnet Station

Connectors to external level and/or temperature instruments. If an AMI

Liquid Helium Level Instrument is purchased with the Model 430

Programmer and magnet system, an LHe level cable will be provided.

Warning

Although the LHe level sensor connector terminals are isolated from

earth ground and therefore touching one terminal is not hazardous,

the voltage between terminals is at a hazardous potential if an AMI

Liquid Helium Level Instrument is connected and energized. The

LHe level sensor pins are designed for use with an AMI LHe sensor

and the wiring for the sensor is to have no live parts which are

accessible. Conductors connected to its terminals must be insulated

from user contact by basic insulation rated for 150 VAC

(Category I).

Note

For maximum noise immunity, use shielded cabling and connect

one end of the shield to the LHe Level / Temp Connector shell.

A.3

Programmer Shunt Terminals

The shunt terminals should be connected so that positive conventional

current flows from the + terminal to the − terminal. Refer to section 2 on

page 11 for a detailed description of the system interconnections.

Warning

Exercise caution near the shunt terminals when operating a magnet.

Metallic objects shorted across the shunt terminals may conduct

large DC currents which are capable of melting the object and

causing severe burns.

Caution

Do not overtighten the nuts on the shunt terminals of the Model 430

Programmer (refer to specifications in table on page 8).

Overtightening can result in damage to the terminals.