2 coil constant, Table 3-4, Maximum recommended stability setting changes – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 68: Operation

Advertising
background image

50

Rev. 5

Operation

Setup Menu : Load

reason, small changes in Stability Setting have a large effect on

stability as the Stability Setting value approaches 100%. Changing

the Stability Setting from 99.9% to 99.8% changes the gain

multiplier from 0.1% to 0.2% (changing the gain multiplier by a

factor of 2, a 100% increase in the gain multiplier). Note, however,

that the same 0.1% change in Stability Setting from 90% to 89.9%

only changes the gain multiplier from 10% to 10.1% (changing the

gain multiplier by a factor of 1.01, a 1% change in gain multiplier).

Likewise a 0.1% change in Stability Setting from 50% to 49.9% only

changes the gain multiplier from 50% to 50.1% (changing the gain

multiplier by a factor of only 1.002, a 0.2% increase in the gain

multiplier).

What this means is that if the Stability Setting is being adjusted to

experimentally determine its optimum setting (using the graph

above as a starting point), no greater than 0.1% changes should be

made above 98%, no greater than 0.2% changes should be made

between 94% and 98%, no greater than 0.5% changes should be

made between 84% and 94%, no greater than 2% changes should be

made between 44% and 84% and no greater than 5% changes

should be made below 44%. See the summary in the table below.

3.10.2.2

Coil Constant

The coil constant is a scaling factor which converts the current to

kilogauss (kG) or tesla (T). It is also often referred to as the field-to-

current ratio. The coil constant is specified in kilogauss/ampere or

tesla/ampere. The value can be set by using either the numeric

keypad per section 3.3 on page 25 or the fine adjust knob

(section 3.4 on page 26). If the coil constant value is 0.0 kG/A (or 0.0

T/A), then no conversion from amperes to kilogauss or tesla is

Table 3-4.

Maximum Recommended Stability Setting Changes

Stability Setting Range

Maximum Recommended

Stability Setting Change

98% to 100%

0.1%

94% to 98%

0.2%

84% to 94%

0.5%

44% to 84%

2%

0% to 44%

5%

+0.00 A — Coil Constant (kG/A)
+0.50 Vs

0.90000

Advertising