5.2.10 Cannot place the magnet in persistent mode.

Ensure there is adequate LHe level in the cryostat to allow the

persistent switch to cool to the superconducting state.

5.2.11 Cannot bring the magnet out of persistent mode.

1. If a PSwitch Error was indicated when the

PERSIST. SWITCH

CONTROL

key was used to turn on the persistent switch heater

current, then there is a problem with the wiring to the persistent

switch heater. Check the continuity between the persistent switch

heater power supply output pins at the rear panel

MAGNET

STATION CONNECTORS

and the connectors on the magnet

support stand top plate. Refer to Table A-1 on page 145.

2. Verify that the output of the persistent switch heater is set to the

appropriate value. Refer to section 3.10.2.8 and Figure 3-13 on page

73.

3. Ensure that there is sufficient time for the switch to warm before

the power supply current is changed. Increase the persistent switch

heating time if needed. Refer to section 3.10.2.9.

5.2.12 The magnet quenches for no apparent reason

1. Ensure the magnet is not being charged at a ramp rate exceeding

the capabilities of the magnet. Exceeding the designed rate for

ramping the magnet may cause a quench or it may turn on

protective diodes on the magnet which may appear very similar to a

quench.

2. Ensure there is adequate LHe level in the cryostat. For systems

operating at less then 4.2K, ensure the magnet is cooled to the

temperature specified by the magnet manufacturer.

3. For conduction-cooled magnets, ensure the magnet temperature is

proper and in accordance with the magnet manufacturer’s

specifications.

4. Disable the Model 430 Programmer quench detection feature (see

section 3.15.2 on page 88) if you suspect the Model 430 Programmer

is falsely indicating a quench condition.

5.2.13 Cannot lower the magnet field

1. Ensure the magnet is not in the persistent mode. Refer to section

3.13.3 for the procedure to remove a magnet from the persistent

mode of operation.