Figure 4-1. the model 430 programmer status system, 1 status byte register, Figure 4-1 the model 430 programmer status system – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 119: Remote interface reference, Rev. 5 101, Scpi status system

Advertising
background image

Rev. 5

101

Remote Interface Reference

SCPI Status System

An event register latches various events. Events are not buffered, therefore

once a bit is set in the event register, further occurrences of that event are

ignored. Once a bit is set in an event register, the bit remains set until the

register is read (

*ESR?

) or a

*CLS

(clear status) command is issued. A

query of an event register returns a decimal value in the appropriate

output buffer which corresponds to the binary-weighted sum of all bits set

in the register.

An enable register (or bitmask) defines which bits in an event register are

reported to the Status Byte register group. An enable register can be both

written and queried. The

*CLS

(clear status) command does not clear an

enable register. To enable or disable bits in an enable register, write a

decimal value which corresponds to the binary-weighted sum of the bits

you wish reported to the Status Byte register.

4.2.2.1

Status Byte Register

The Status Byte register group reports conditions from the Standard Event
register or output buffers. Data in the output buffer is immediately reported in
the “Serial Message Available” bit (bit 3). Clearing a bit in the Standard Event
register will update the corresponding bit in the Status Byte register,
according to the Standard Event enable register. Reading the pending

0

1

2

3

4

5

6

7

<1>

<2>

<4>

<8>

<16>

<32>

N/A

<128>

*SRE <

value>

*SRE?

Serial Poll

*STB?

C

EN

Summary Bit

"OR"

Status Byte Register

0

1

2

3

4

5

6

7

<1>

<2>

<4>

<8>

<16>

<32>

<64>

<128>

*ESE <

value>

*ESE?

*ESR?

EV

EN

"OR"

Standard Event Register

Operation Complete

Query Error

Device Error

Execution Error

Command Error

Power On

Quench Condition

Serial Output Buffer

NOTES:
C = Condition Register
EV = Event Register
EN = Enable Register

Figure 4-1.

The Model 430 Programmer Status System.

Advertising