Remote interface reference – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 137

Advertising
background image

Rev. 5

119

Remote Interface Reference

SETUP Configuration Commands and Queries

CONFigure:LOCK:COILconst {0|1}

Specifies whether the Coil Constant value (under the Load submenu) is

locked or unlocked. Sending “0” unlocks. A “1” locks. “0” is the default

value.

LOCK:VOLTage:COILconst?

Returns “0” for Coil Constant value (under the Load submenu) unlocked,

or “1” for locked.

CONFigure:LOCK:CURRent:LIMit {0|1}

Specifies whether the Current Limit value (under the Load submenu) is

locked or unlocked. Sending “0” unlocks. A “1” locks. “0” is the default

value.

LOCK:CURRent:LIMit?

Returns “0” for Current Limit value (under the Load submenu) unlocked,

or “1” for locked.

CONFigure:LOCK:PSwitch:SETtings {0|1}

Specifies whether persistent switch settings are locked or unlocked.

Persistent switch settings protected by this setting (all under the Load

submenu) are: the PSwitch Installed picklist value, the PSwitch Current

value, the PSwitch Heated Time value, PSwitch Current Detect, PSwitch

Cooled time, PSw P/S Ramp Rate, and PSwitch Cooling Gain. Sending “0”

unlocks. A “1” locks. “0” is the default value.

LOCK:PSwitch:SETtings?

Returns “0” for persistent switch settings unlocked, or “1” for locked.

CONFigure:LOCK:QUench:DETect {0|1}

Specifies whether the Enable Quench Detect picklist value (under the

Load submenu) is locked or unlocked. Sending “0” unlocks. A “1” locks. “0”

is the default value.

LOCK:QUench:DETect?

Returns “0” for Enable Quench Detect picklist value (under the Load

submenu) unlocked, or “1” for locked.

CONFigure:LOCK:ABsorber {0|1}

Specifies whether the Energy Absorber Present picklist value (under the

Load submenu) is locked or unlocked. Sending “0” unlocks. A “1” locks. “0”

is the default value.