Operation – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 67

Advertising
background image

Rev. 5

49

Operation

Setup Menu : Load

by using either the numeric keypad per section 3.3 on page 25 or

the fine adjust knob (section 3.4 on page 26). The valid range is

from 0.0 to 100.0%. The default value is 0.0% unless preset by AMI

to match a specific superconducting magnet.

Superconducting magnets without a persistent switch require

1

a

specific Model 430 Programmer stability setting based on the

magnet inductance as follows:

For magnet inductance <= 100 Henries (H):

Stability Setting = (100 - H)

For magnet inductance > 100 Henries:

Stability Setting = 0

The graph below may be used as a guide to set the stability setting

for magnets with a persistent switch installed and inductance of

less than 3 henries. Magnets with an inductance of greater than 3

henries that have a persistent switch installed should operate with

a stability setting of 0.0%.

The Model 430 Programmer internal control loop gain is

proportional to the multiplier (100% – [Stability Setting]), except

that for a Stability of 100% the multiplier is set to a low non-zero

value suitable for controlling current in a short circuit. For this

1. Effective with Model 430 firmware version 1.62.

y

g

g

0

10

20

30

40

50

60

70

80

90

100

0

1

2

3

4

5

6

7

Inductance (H)

Stab

ility Settin

g

(%)

Figure 3-9.

Stability Setting vs. Magnet (with PSwitch) Inductance

Model 430 Stability Setting vs. Magnet Inductance

(for Magnets with Persistent Switch)