Operation

Persistent Switch Control : Entering Persistent Mode

Note

The magnet voltage (Vm) is monitored during the power supply

ramp to zero. If the magnet voltage exceeds 0.5 V during this ramp,

the ramp is paused and the Model 430 Programmer beeps to

indicate the persistent switch did not transition to the

superconducting state properly. If this error occurs, the Model 430

will ramp the current back to the value when the persistent switch

was cooled,

and then heat the switch:

After the persistent switch heated time has elapsed, the display

will indicate the persistent mode transition malfunction:

After pressing

ENTER, the Model 430 Programmer will revert to

the default field/current display.

5. After the power supply is finished ramping to zero, the following

screen will be displayed:

+45.39 A

Mode: Power Supply Ramping

+3.44 Vs

to match magnet field

+50.00 A

Mode: Heating Switch

+3.50 Vs

***** PSW Lock Error *****

+50.00 A -

Mode: Paused w/PSW error

+3.50 Vs

Press ENTER to continue

+50.00 A

P

Mode: Paused

+3.50 Vs

PSwitch Heater: ON

+0.00 A -

Magnet in Persistent Mode

+0.00 Vm

Press PER.SW.CTRL. to exit