Remote interface reference – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 131

Advertising
background image

Rev. 5

113

Remote Interface Reference

SETUP Configuration Commands and Queries

COILconst?

Returns the coil constant setting in kG/A or T/A per the selected field

units.

CONFigure:CURRent:LIMit <

current (A)

>

Sets the Current Limit in amperes. The Current Limit is the largest

magnitude operating current allowed during any ramping mode. For four-

quadrant power supplies, the Current Limit functions as both a positive

and negative current limit.

CURRent:LIMit?

Returns the Current Limit in amperes.

CONFigure:CURRent:RATING <

current (A)

>

Sets the magnet current rating in amperes.

CURRent:RATING?

Returns the current magnet rating in amperes.

CONFigure:PSwitch {0|1}

PSwitch:INSTalled?

“0” indicates that a persistent switch is not installed on the connected

superconducting magnet. “1” indicates that a persistent switch is installed.

If a persistent switch is installed, the persistent switch heated current and

time should be specified. The default value is “1”.

CONFigure:PSwitch:CURRent <

current (mA)

>

Sets the persistent switch heater current in mA.

PSwitch:AUTODetect?

Executes the auto-detection algorithm (Refer to section 3.10.2.7 on

page 54) and returns the appropriate persistent switch heater current in

mA. Note that after this value is returned, it can be entered into the Model

430 Programmer using the

CONFigure:PSwitch:CURRent

command.

PSwitch:CURRent?

Returns the persistent switch heater current setting in mA.

CONFigure:PSwitch:HeatTIME <

time (seconds)

>

Sets the time required in seconds for the persistent switch to become

resistive after the persistent switch heater has been activated.

PSwitch:HeatTIME?

Returns the persistent switch heated time in seconds.