Figure 3-1. default display, 1 field / current display, Figure 3-1 default display – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 41: Field). refer to sections 3.2.1 an, Operation

Advertising
background image

Rev. 5

23

Operation

Default Display : Field / Current

display areas — the Field / Current Display area, the Voltage Display

area, the Status Indicator area and the Main Display area.

3.2.1

Field / Current Display

The field / current display indicates either the field strength or current

1

.

This is always displayed in the upper left corner of the display (see Figure

3-1), regardless of what else is being displayed on the Model 430

Programmer display. The parameter displayed (field or current) is toggled

by pressing

SHIFT

followed by

FIELD <> CURRENT

. Thus, if field strength

is being displayed, pressing

SHIFT

followed by

FIELD <> CURRENT

will

cause the current to be displayed; conversely, if current is being displayed,

pressing

SHIFT

followed by

FIELD <> CURRENT

will cause the field

strength to be displayed. Operating current is always displayed in

amperes (A). Operating field strength may be displayed in kilogauss (kG)

or tesla (T) if a coil constant has been specified in the setup

2

. If field

strength is being displayed, the units (kG or T) in which it is displayed can

be toggled by pressing

SHIFT

followed by

FIELD UNITS

.

Note

Note that the displayed field strength is not directly measured, but

rather is calculated by multiplying the coil constant entered in the

setup menu by the measured current flow of the Model 430 power

supply system.

1. The value is always displayed in current (A) when an installed persistent switch is in

the

cooled state

since the value represents power supply current only, independent

of magnet current/field.

2. Refer to section 3.10.2.2 on page 50.

+50.00 A — Status: Holding

+0.50 Vs

PSwitch Heater: ON

+50.00 A —

-10

0Vm

+10

+1.50 V s

|

''''|''''|''''|'''''|

Field / Current Display.

Default display showing ramp mode and persistent switch heater status:

Default display showing voltmeter:

Figure 3-1.

Default Display.

Voltage Display.

Status Indicator.

Main Display.

Advertising