Table 4-4, Return values and meanings for, Supply:m – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 130: Remote interface reference

Advertising
background image

112

Rev. 5

Remote Interface Reference

SETUP Configuration Commands and Queries

SUPPly:CURRent:MINimum?

Returns the minimum output current capacity of the power supply in

amperes. This value can be configured only via front panel operation using

the Supply submenu and is set automatically when a preset supply type is

selected.

SUPPly:CURRent:MAXimum?

Returns the maximum output current capacity of the power supply in

amperes. This value can be configured only via front panel operation using

the Supply submenu and is set automatically when a preset supply type is

selected.

SUPPly:MODE?

Returns an integer value corresponding to the voltage output mode

according to the table below:

This value can be configured only via front panel operation using the

Supply submenu and is set automatically when a preset supply type is

selected.

CONFigure:STABility <

percent

>

Sets the stability setting in percent.

STABility?

Returns the stability setting in percent.

CONFigure:COILconst <

value (kG/A, T/A

>

Sets the coil constant (also referred to as the field-to-current ratio) per the

selected field units. The coil constant must be set to a non-zero, positive

value in order to command or query the Model 430 Programmer in units of

field.

Table 4-4.

Return Values and Meanings for

SUPPly:MODE?

Query

Return

Value

Meaning

0

+0.000 to +5.000

1

+0.000 to +10.000

2

-5.000 to +5.000

3

-10.000 to +10.000

4

+0.000 to -5.000

5

+0.000 to +8.000