74

Rev. 5

Operation

Ramping Functions

If your magnet, Model 430 Programmer, and power supply were purchased

as a system from AMI, the setup menus are preset by AMI to match the

magnet purchased.

Table 3-5 provides a summary of the Model 430 Programmer setup

parameters for this example.

3.12 Ramping Functions

The ramping functions are used to control charging of the superconducting

load. The Model 430 Programmer allows piecewise-linear charging profiles

Table 3-5.

Example Setup Configuration

Parameter

Setting

Select Supply

AMI 4Q06125PS

Stability Setting (%)

0.0

Coil Constant

a

(kG/A)

a. Also referred to as the Field-to-Current Ratio. Obtained by

dividing the rated field by the rated current if not explicitly
stated.

0.934

Current Limit (A)

85.600

PSwitch Installed

YES

PSwitch Current (mA)

41.0

PSwitch Heated Time (sec)

20

PSwitch Cooled Time (sec)

20

PSwitch P/S Ramp Rate (A/sec)

10

Enable Quench Detect

YES

Energy Absorber Present

NO

Voltage Limit (V)

4.100

b

b. Value is the 2.1 V charge rate plus allowances for power

lead drop at the rated current.

Ramp Rate (A/sec)

0.2165

c

c. Value is obtained by dividing the magnet charging voltage

(V) by the magnet inductance (H).

Magnet Current Rating (A)

85.600

PSwitch Cooling Gain (%)

0.0%

Extern Rampdown Enabled

YES