Operation – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 97

Advertising
background image

Rev. 5

79

Operation

Persistent Switch Control : Entering Persistent Mode

2. The Model 430 Programmer must be in either the HOLDING or

PAUSED mode at the target field or current.

3. The Model 430 Programmer must be at the default field/current

display.

1

4. Press the

PERSIST. SWITCH CONTROL

key to turn off the

persistent switch heater current and automatically ramp the power

supply to zero current:

a. After the

PERSIST. SWITCH CONTROL

key is pressed, the

Model 430 Programmer requests that the

ENTER

key be

pressed as a confirmation that the magnet should be placed in

persistent mode.

2

b. When

ENTER

is pressed, the persistent switch is cooled for the

preset persistent switch cooling time (set by the PSwitch Cooled

Time variable

3

). The display indicates that the persistent

switch is being cooled and indicates the number of seconds (4 in

this example) remaining in the cooling cycle.

c. When the cooled time is complete, the green

MAGNET IN

PERSISTENT MODE

LED will illuminate

4

and the power

supply will ramp to zero at the PSw P/S Ramp Rate value

5

.

1. Refer to section 3.2.1 on page 23.
2. Pressing the

ESC

ape key will terminate the command and return the Model 430

Programmer to the default screen.

3. Refer to section 3.10.2.10 on page 56.
4. The threshold for this LED is 100 mA of magnet current.
5. Refer to section 3.10.2.11 on page 56.

+50.00 A -

Press ENTER to begin

+3.50 Vs

Persistent Mode

+50.00 A

Mode: Cooling Switch (4)

+3.50 Vs

PSwitch Heater: OFF

+42.89 A

Mode: Power Supply ramping

+3.38 Vs

to zero current