2 Installation

Warning

Before energizing the equipment, the earth ground of the power

receptacle must be verified to be at earth potential and able to carry

the rated current of the power circuit. Using extension cords should

be avoided; however, if one must be used, ensure the ground

conductor is intact and capable of carrying the rated current.

In the event that the ground path becomes less than sufficient to

carry the rated current of the power circuit, the equipment should be

disconnected from power, labeled as unsafe, and removed from place

of operation.

Do not operate this equipment in the presence of flammable gases.

Doing so could result in a life-threatening explosion.

Do not modify this equipment in any way. If component replacement

is required, return the equipment to AMI facilities as described in

the Troubleshooting section of this manual.

If used in a manner not specified in this manual, the protection

provided by the design, manufacture and documentation of the

system may be impaired.

2.1 Inspecting and Unpacking

Carefully remove the equipment, interconnecting cabling, and

documentation CD (and/or printed material) from the shipping carton, and

remove all packaging material.

Note

If there is any shipping damage, save all packing material and

contact the shipping representative to file a damage claim. Do not

return to AMI unless prior authorization has been received.

2.2 Power Supply System Mounting

If the system is to be used on a table top, place the equipment on a flat,

secure surface.