Remote Interface Reference

Ramp Configuration Commands and Queries

section 3.12 for additional information on determining ramp rates. Also

included are queries for collecting the magnet field, current, voltage, and

inductance.

CONFigure:VOLTage:LIMit <

voltage (V)

>

Sets the ramping Voltage Limit in volts. The ramping Voltage Limit may

not exceed the maximum output voltage of the power supply.

VOLTage:LIMit?

Returns the ramping Voltage Limit in volts.

CONFigure:CURRent:TARGet <

current (A)

>

Sets the target current in amperes.

CURRent:TARGet?

Returns the target current setting in amperes.

CONFigure:FIELD:TARGet <

field (kG, T)

>

Sets the target field in units of kilogauss or tesla, per the selected field

units. This command requires that a coil constant be defined, otherwise an

error is generated.

FIELD:TARGet?

Returns the target field setting in units of kilogauss or tesla, per the

selected field units. This query requires that a coil constant be defined,

otherwise an error is generated.

CONFigure:RAMP:RATE:SEGments <

value

>

Sets the number of ramp segments (see section 3.7.1 for details of the use

of ramp segments).

RAMP:RATE:SEGments?

Returns the number of ramp segments.

CONFigure:RAMP:RATE:CURRent <

segment>,<rate (A/s, A/min)>,

<upper bound (A)

>

Sets the ramp rate for the specified segment (values of 1 through the

defined number of ramp segments are valid) in units of A/sec or A/min (per

the selected ramp rate units), and defines the current upper bound for that

segment in amperes (see section 3.7.1 for details of the use of ramp

segments).

RAMP:RATE:CURRent:<

segment>

?

Returns the ramp rate setting for the specified segment (values of 1

through the defined number of ramp segments are valid) in units of A/sec