Appendix

Auxiliary LHe Level/Temperature Connectors

Note

For maximum noise immunity, use shielded cabling and connect

one end of the shield to the Magnet Station Connector shell.

The connectors provide an interface for connecting a single integrated

instrumentation cable from the magnet support stand to the Model 430

Programmer. The Model 430 Programmer can then be used to distribute

the signals to the appropriate instruments or data acquisition systems.

The LHe level and temperature sensor signals are also internally routed to

the LHe Level / Temp Connectors.

If the Model 430 Programmer is purchased as part of a magnet system, a

Magnet Station Connector instrumentation cable will be provided with the

system.

A.2

LHe Level / Temp Connectors

The two 9-pin D-sub male LHe Level / Temp Connectors are identically

wired and connected pin-for-pin internally.

Table A-2.

LHe Level / Temp Connectors Pin Definitions

Pin

Function

1

LHe Sensor I+ (Red)

2

Temperature Sensor I+ (Red)

3

Temperature Sensor V

(Yellow)

4

Temperature Sensor I

(Black)

5

Temperature Sensor V+ (Blue)

6

LHe Sensor V

(Yellow)

7

LHe Sensor I

(Black)

8

LHe Sensor V+ (Blue)

9

not used