Table 3-6. ramp modes and descriptions, 2 manual ramping, 3 automatic ramping – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 94: Table 3-6, Ramp modes and descriptions, E 3-6, Operation

Advertising
background image

76

Rev. 5

Operation

Ramping Functions : Manual Ramping

Limit and ramp rate(s) are applicable to both manual and automatic

ramping.

3.12.2 Manual

Ramping

The

INCR. FIELD

and

DECR. FIELD

SHIFT-key functions control manual

ramping. Manual ramping ramps field/current up or down at the defined

ramp rate(s). See section 3.7.4 on page 37 and section 3.7.6 on page 38 for

details regarding the use of these SHIFT-key functions.

3.12.3 Automatic

Ramping

Automatic ramping differs from manual ramping in that the Model 430

Programmer automatically performs ramping in the appropriate direction

to achieve the value of the target field/current setting. To use automatic

ramping, enter the target field/current with which ramping is desired

1

. If

ramping is not PAUSED, ramping to the target field/current begins

immediately. If ramping is PAUSED, ramping to the target field/current

will begin when the

RAMP / PAUSE

key is pressed to take the Model 430

1. Refer to section 3.7.1 on page 33 and section 3.7.2 on page 36.

Table 3-6.

Ramp modes and descriptions.

Mode

Description

Ramping

Automatic ramping to the target field/current

a

is in progress.

Holding

The target field/current has been achieved and is being
maintained.

Paused

Ramping is suspended at the field/current achieved at the time
the PAUSED mode was entered.

Manual

Ramping is being controlled by the manual control (

INCR.

FIELD

and

DECR. FIELD

) SHIFT-key functions available on

the front panel.

Zeroing

Current

RAMP TO ZERO

is active, and the Model 430 Programmer is

ramping current to 0 A.

Zero Current

RAMP TO ZERO

is still active, and the current is less than

0.1% of

I

max

.

Heating

Switch

The persistent switch heater has been activated. Ramping is
disabled during the persistent switch heating period.

Cooling

Switch

The persistent switch heater has been deactivated. Ramping is
disabled during the persistent switch cooling period.

a. The target field/current setting is discussed in section 3.6.2.

1. Refer to section 3.6.2 on page 31.

Advertising