1 low-current, high-resolution 4-quadrant supply, Installation – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 32

Advertising
background image

14

Rev.

5

Installation

Low-Current, High-Resolution 4-Quadrant Supply

2.5.1 Low-Current, High-Resolution 4-Quadrant Supply

For this four-quadrant low-current power supply system, components

include a Model 430 Programmer, a Model 4Q1010PS low-current four-

quadrant power supply, and associated interconnecting cabling. Figure 2-2

on page 15 and Figure 2-1 illustrate the interconnects for the Model

4Q1010PS-430 low-current integrated power supply system.

Note

Due to continuous discharge power limitations present in the BOP

series supplies, for maximum safety AMI limits both the charging

and discharging voltage, via the Model 430 Programmer, to a

maximum of 10 volts.

Caution

Do not overtighten the hardware on the interconnection terminals

(refer to specifications table on page 8 for torque limits).

Overtightening can result in damage to the terminals.

Refer to Figure 2-2 on page 15. Ensure the cabling is connected in the

following manner:

a. Connect the

OUT

power supply terminal (1) to the positive magnet

current lead (2).

Figure 2-1.

Typical Model 4Q1010PS-430 System Rack Interconnections

Advertising