Power-Up Procedure

appears to lag, then decrease the stability setting until the system is

responsive. If the current appears to oscillate, increase the stability setting

until the oscillations are damped.

Note

If you have purchased a superconducting magnet with the Model

430 Programmer, AMI will normally provide a recommended

stability setting for optimal operation of the magnet system. If you

operate the Model 430 Programmer with a different load, be sure to

restore the stability setting to the recommended value when the

superconducting magnet is reconnected.

The stability setting is essentially manual control of the gain of an

integrator present in the control logic of the Model 430 Programmer.

Increasing the stability setting decreases the gain of the integrator.

2.9 Power-Up and Test Procedure

It is important to verify that the magnet system has been properly

connected before the superconducting magnet is energized. This is

especially recommended if the system is to be controlled via a computer

since this setup will allow software debugging without the potential for

damage to the magnet. The following procedures will assist the user in the

verifying key system components.

1. Using the appropriate diagram from section 2.5 as a guide, verify

all system components are connected as shown. If there is any

doubt as to the correct connection of a component, contact an AMI

Technical Support Representative. The user may be required to

properly make a few connections between the various system

components which were disconnected to facilitate packing and

shipping.

2. Temporarily place a short across the magnet current terminals.

Often this is most easily accomplished by unfastening the heavy

cables from the magnet current leads and fastening them together.

1

This will allow rudimentary power supply checks without

energizing the superconducting magnet.

3. Energize the Model 430 Programmer by placing the power switch

in the

I (ON) position.

1. If the system shipped with CamLoc quick-disconnect connectors, they may be

quickly disconnected from the magnet leads and connected together.

Advertising