2 query errors, Remote interface reference – American Magnetics 4Q1010PS-430 Integrated Power Supply System User Manual

Page 149

Advertising
background image

Rev. 5

131

Remote Interface Reference

Error Messages

-103,”Non-boolean argument”

The command required a parameter in the form of 0 or 1. No other form of

the parameter is allowed.

-104,”Missing parameter”

The command required at least one argument which was not found before

the termination character(s).

-105,”Out of range”

At least one of the parameter values received was out of the valid range.

Refer to the summary of valid ranges for the Model 430 Programmer

settings on page 90. Be sure to note the field units and ramp units settings

and check any unit conversions.

-106,”Undefined coil const”

The user attempted to invoke a command with units of field without first

setting a value for the coil constant. The coil constant must be a non-zero,

positive value.

-107,”No switch installed”

The user attempted to activate the persistent switch heater when no

switch is installed. Before activating the persistent switch heater, the user

must indicate a switch is installed and set the switch current and heating

time (see page 54).

-

108,“Not Ramping”

The command that was issued requires the Model 430 Programmer to be

ramping for the duration of command processing, and the Model 430

Programmer was either not ramping when the command was issued, or

stopped ramping before the command processing was completed.

4.6.2

Query Errors

-201,”Unrecognized query”

The query string sent (identified as a query by a

?

) was not identified as

valid. Check the query string for invalid characters or separators, syntax

errors, or for errors in the mnemonics. Spaces are not allowed before or

after colon separators.

-202,”Undefined coil const”

The user attempted to invoke a query with units of field without first

setting a value for the coil constant. The coil constant must be a non-zero,

positive value.